| SKU | HI83300-01 |
| Name | Multiparameter Benchtop Photometer and pH meter – HI83300 |
| pH Range | Photometer: 6.5 to 8.5 pH
pH electrode: -2.00 to 16.00 pH |
| pH Resolution | Photometer: 0.1 pH
pH electrode: 0.1 pH |
| pH Accuracy | Photometer: ±0.1 pH
pH electrode: ±0.01 pH |
| pH Calibration | Automatic one or two point calibration with one set of standard buffers available (4.01, 6.86, 7.01, 9.18, 10.01) |
| pH Temperature Compensation | Automatic (-5.0 to 100.0 ℃; 23.0 to 212.0 °F); limits reduced based on the pH electrode used |
| pH CAL Check (electrode diagnostics) | clean electrode and check buffer/check probe displayed during calibration |
| pH Method | Photometer: phenol red |
| pH-mV Range | ±1000 mV |
| pH-mV Resolution | 0.1 mV |
| pH-mV Accuracy | ±0.2 mV |
| Dissolved Oxygen Range | 0.0 to 10.0 mg/L (as O2) |
| Dissolved Oxygen Resolution | 0.1 mg/L |
| Dissolved Oxygen Accuracy | ±0.4 mg/L ±3% of reading |
| Dissolved Oxygen Measurement Method | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Azide modified Winkler method |
| Absorbance Range | 0.000 to 4.000 Abs |
| Absorbance Resolution | 0.001 Abs |
| Absorbance Accuracy | ±0.003Abs @ 1.000 Abs |
| Alkalinity Range | Freshwater: 0 to 500 mg/L (as CaCO3)
Seawater: 0 to 500 mg/L (as CaCO3) |
| Alkalinity Resolution | 1 mg/L |
| Alkalinity Accuracy | ±5 mg/L ±5% of reading |
| Alkalinity Method | Colorimetric method |
| Aluminum Range | 0.00 to 1.00 mg/L (as Al3+) |
| Aluminum Resolution | 0.01 mg/L |
| Aluminum Accuracy | ±0.04 mg/L ±4% of reading |
| Aluminum Method | Adaptation of the aluminon method |
| Ammonia Range | Low Range: 0.00 to 3.00 mg/L (as NH3-N)
Medium Range: 0.00 to 10.00 mg/L (as NH3-N)
High Range: 0.0 to 100.0 mg/L (as NH3-N) |
| Ammonia Resolution | 0.01 mg/L; 0.1 mg/L |
| Ammonia Accuracy | Low Range: ±0.04 mg/L ±4% of reading
Medium Range: ±0.05 mg/L ±5% of reading
High range: ±0.5 mg/L ±5% of reading at 25 °C |
| Ammonia Method | Adaptation of the ASTM Manual of Water and Environmental Technology, D1426-92, Nessler method |
| Anionic Surfactants Range | 0.00 to 3.50 mg/L (as SDBS) |
| Anionic Surfactants Resolution | 0.01 mg/L |
| Anionic Surfactants Accuracy | ±0.04 mg/L ±3% of reading |
| Anionic Surfactants Method | Adaptation of the USEPA method 425.1 and Standard Methods for the Examination of Water and Wastewater, 20th edition, 5540C, Anionic Surfactants as MBAS. |
| Bromine Range | 0.00 to 8.00 mg/L (as Br2) |
| Bromine Resolution | 0.01 mg/L |
| Bromine Accuracy | ±0.08 mg/L ±3% of reading |
| Bromine Method | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, DPD method. |
| Calcium Range | Freshwater: 0 to 400 mg/L (as Ca2+)
Seawater: 200 to 600 mg/L (as Ca2+) |
| Calcium Resolution | 1 mg/L |
| Calcium Accuracy | Freshwater: ±10 mg/L ±5% of reading
Seawater: ±6% of reading |
| Calcium Method | Freshwater: Adaptation of the Oxalate method
Seawater: Adaptation of the Zincon method |
| Chloride Range | 0.0 to 20.0 mg/L (as Cl⁻) |
| Chloride Resolution | 0.1 mg/L |
| Chloride Accuracy | ±0.5 mg/L ±6% of reading at 25 °C |
| Chlorine Dioxide Range | 0.00 to 2.00 mg/L (as ClO2) |
| Chlorine Dioxide Resolution | 0.01 mg/L |
| Chlorine Dioxide Accuracy | ±0.10 mg/L ±5% of reading |
| Chlorine Dioxide Method | Adaptation of the Chlorophenol Red method. |
| Free Chlorine Range | 0.00 to 5.00 mg/L (as Cl2)
Ultra Low Range: 0.000 to 0.500 mg/L (as Cl2) |
| Free Chlorine Resolution | 0.01 mg/L
Ultra Low Range: 0.001 mg/L |
| Free Chlorine Accuracy | ±0.03 mg/L ±3% of reading
Ultra Low Range: ±0.020 mg/L ±3% of reading |
| Total Chlorine Range | 0.00 to 5.00 mg/L (as Cl2)
Ultra Low Range: 0.000 to 0.500 mg/L (as Cl2)
Ultra High Range: 0 to 500 mg/L (as Cl2) |
| Total Chlorine Resolution | 0.01 mg/L
Ultra Low Range: 0.001 mg/L
Ultra High Range:1 mg/L |
| Total Chlorine Accuracy | ±0.03 mg/L ±3% of reading
Ultra Low Range: ±0.020 mg/L ±3% of reading
Ultra High Range: ±3 mg/L ±3% of reading |
| Chlorine Method | Adaptation of the EPA 330.5 DPD method
Free Chlorine (ULR) & Total Chlorine (UHR): Adaptation of the Standard Methods for Examination of Water and Wastewater, 20th edition, 4500-Cl |
| Chromium, Hexavalent Range | Low Range: 0 to 300 µg/L (as Cr6+)
High Range: 0 to 1000 µg/L (as Cr6+) |
| Chromium, Hexavalent Resolution | 1 µg/L |
| Chromium, Hexavalent Accuracy | Low Range: ±1 µg/L ±4% of reading
High Range: ±5 µg/L ±4% of reading |
| Chromium, Hexavalent Method | Adaptation of the ASTM Manual of Water and Environmental Technology, D1687-92, Diphenylcarbohydrazide method. |
| Color, Water Range | 0 to 500 PCU (Platinum Cobalt Units) |
| Color, Water Resolution | 1 PCU |
| Color, Water Accuracy | ±10 PCU ±5% of reading |
| Color, Water Method | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Colorimetric Platinum Cobalt method. |
| Copper Range | Low Range: 0.000 to 1.500 mg/L (as Cu2+)
High range: 0.00 to 5.00 mg/L (as Cu2+) |
| Copper Resolution | 0.001 mg/L; 0.01 mg/L |
| Copper Accuracy | Low Range: ±0.01 mg/L ±5% of reading
High Range ±0.02 mg/L ±4% of reading |
| Copper Method | Adaptation of the EPA bicinchoninate method |
| Cyanuric Acid Range | 0 to 80 mg/L (as CYA) |
| Cyanuric Acid Resolution | 1 mg/L |
| Cyanuric Acid Accuracy | ±1 mg/L ±15% of reading |
| Cyanuric Acid Method | Adaptation of the turbidimetric method |
| Fluoride Range | Low Range: 0.00 to 2.00 mg/L (as F–)
High range: 0.0 to 20.0 mg/L (as F–) |
| Fluoride Resolution | 0.01 mg/L; 0.1 mg/L |
| Fluoride Accuracy | Low Range: ±0.03 mg/L ±3% of reading
High Range: ±0.5 mg/L ±3% of reading |
| Fluoride Method | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, SPADNS method |
| Hardness, Total Range | Low Range: 0 to 250 mg/L (as CaCO3)
Medium Range: 200 to 500 mg/L (as CaCO3)
High Range: 400 to 750 mg/L (as CaCO3) |
| Hardness, Total Resolution | 1 mg/L |
| Hardness, Total Accuracy | Low Range: ±5 mg/L ±4% of reading
Medium Range: ±7 mg/L ±3% of reading
High Range: ±10 mg/L ±2% of reading |
| Hardness, Total Method | Adaptation of the EPA recommended method 130.1 |
| Hardness, Calcium Range | 0.00 to 2.70 mg/L (as CaCO3) |
| Hardness, Calcium Resolution | 0.01 mg/L |
| Hardness, Calcium Accuracy | ±0.11 mg/L ±5% of reading |
| Hardness, Calcium Method | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Calmagite method |
| Hardness, Magnesium Range | 0.00 to 2.00 mg/L (as CaCO3) |
| Hardness, Magnesium Resolution | 0.01 mg/L |
| Hardness, Magnesium Accuracy | ±0.11 mg/L ±5% of reading |
| Hardness, Magnesium Method | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, EDTA Colorimetric method |
| Hydrazine Range | 0 to 400 µg/L (as N2H4) |
| Hydrazine Resolution | 1 µg/L |
| Hydrazine Accuracy | ±4% of full scale reading |
| Hydrazine Method | Adaptation of the ASTM Manual of Water and Environmental Technology, method D1385-88, p-Dimethylaminobenzaldehyde method |
| Iodine Range | 0.0 to 12.5 mg/L (as I2) |
| Iodine Resolution | 0.1 mg/L |
| Iodine Accuracy | ±0.1 mg/L ±5% of reading |
| Iodine Method | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, DPD method |
| Iron Range | Low Range: 0.000 to 1.600 mg/L (as Fe)
High Range: 0.00 to 5.00 mg/L (as Fe) |
| Iron Resolution | 0.001 mg/L; 0.01 mg/L |
| Iron Accuracy | Low Range: ±0.01 mg/L ±8% of reading
High Range: ±0.04 mg/L ±2% of reading |
| Iron Method | Low Range: Adaptation of the TPTZ Method
Adaptation of Standard Methods for the Examination of Water and Wastewater, 3500-Fe B., Phenanthroline Method |
| Magnesium Range | 0 to 150 mg/L (as Mg2+) |
| Magnesium Resolution | 1 mg/L |
| Magnesium Accuracy | ±5 mg/L ±3% of reading |
| Magnesium Method | Adaptation of the Calmagite method |
| Manganese Range | Low Range: 0 to 300 µg/L (as Mn)
High Range: 0.0 to 20.0 (as Mn) |
| Manganese Resolution | 1 µg/L; 0.1 mg/L |
| Manganese Accuracy | Low Range: ±10 µg/L ±3% of reading
High Range: ±0.2 mg/L ±3% of reading |
| Manganese Method | Low Range: Adaptation of the PAN Method
High Range: Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Periodate method |
| Molybdenum Range | 0.0 to 40.0 mg/L (as Mo6+) |
| Molybdenum Resolution | 0.1 mg/L |
| Molybdenum Accuracy | ±0.3 mg/L ±5% of reading |
| Molybdenum Method | Adaptation of the mercaptoacetic acid method |
| Nickel Range | Low Range: 0.000 to 1.000 mg/L (as Ni)
High Range: 0.00 to 7.00 g/L (as Ni) |
| Nickel Resolution | 0.001 mg/L; 0.01 g/L |
| Nickel Accuracy | Low range: ±0.010 mg/L ±7% of reading
High Range: ±0.07 g/L ±4% of reading |
| Nickel Method | Low Range: Adaptation of the PAN method
High Range: Adaptation of the photometric method |
| Nitrate Range | 0.0 to 30.0 mg/L (as NO3–– N) |
| Nitrate Resolution | 0.1 mg/L |
| Nitrate Accuracy | ±0.5 mg/L ±10% of reading |
| Nitrate Method | Adaptation of the cadmium reduction method |
| Nitrite Range | Freshwater
Low Range: 0 to 600 µg/L (as NO2–-N)
High Range: 0 to 150 mg/L (as NO2–)
Seawater Ultra Low Range: 0 to 200 µg/L (as NO2–-N) |
| Nitrite Resolution | Freshwater: 1 µg/L; 1 mg/L
Seawater: 1 µg/L |
| Nitrite Accuracy | Freshwater
Low Range: ±20 µg/L ±4% of reading
High Range: ±4 mg/L ±4% of reading
Seawater
±10 µg/L ±4% of reading |
| Nitrite Method | Adaptation of the EPA Diazotization method 354.1 |
| Oxygen, Scavenger Range | 0 to 1000 µg/L (as DEHA)
0.00 to 1.50 mg/L (as Carbohydrazide)
0.00 to 2.50 mg/L (as Hydroquinone)
0.00 to 4.50 mg/L (as ISO-ascorbic acid) |
| Oxygen, Scavenger Resolution | 1 µg/L (DEHA); 0.01 mg/L |
| Oxygen, Scavenger Accuracy | ±5 µg/L ±5% of reading (DEHA) |
| Oxygen, Scavenger Method | Adaptation of the iron reduction method |
| Ozone Range | 0.00 to 2.00 mg/L (as O3) |
| Ozone Resolution | 0.01 mg/L |
| Ozone Accuracy | ±0.02 mg/L ±3% of reading |
| Ozone Method | Colorimetric DPD Method |
| Phosphate Range | Freshwater
Low Range: 0.00 to 2.50 mg/L (as PO43-)
High range: 0.0 to 30.0 mg/L (as PO43-)
Seawater Ultra Low Range: 0 to 200 µg/L (as P) |
| Phosphate Resolution | Freshwater: 0.01 mg/L; 0.1 mg/L
Seawater: 1 µg/L |
| Phosphate Accuracy | Freshwater
Low Range: ±0.04 mg/L ±4% of reading
High Range: ±1 mg/L ±4% of reading
Seawater Ultra Low Range: ±5 µg/L ±5% of reading |
| Phosphate Method | Freshwater Low Range: Adaptation of the Ascorbic Acid method
Freshwater High Range and Seawater Ultra Low Range: Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Amino Acid method |
| Potassium Range | 0.0 to 20.0 mg/L (as K) |
| Potassium Resolution | 0.1 mg/L |
| Potassium Accuracy | ±3.0 mg/L ±7% of reading |
| Potassium Method | Adaptation of the Turbidimetric Tetraphenylborate method |
| Silica Range | Low Range: 0.00 to 2.00 mg/L (as SiO2)
High Range: 0 to 200 mg/L (as SiO2) |
| Silica Resolution | 0.01 mg/L; 1 mg/L |
| Silica Accuracy | Low Range: ±0.03 mg/L ±3% of reading
High Range: ±1 mg/L ±5% of reading |
| Silica Method | Low Range: Adaptation of the ASTM Manual of Water and Environmental Technology, D859, Heteropoly Molybdenum Blue method
High Range: Adaptation of the USEPA Method 370.1 and Standard Method 4500-SiO2 |
| Silver Range | 0.000 to 1.000 mg/L (as Ag) |
| Silver Resolution | 0.001 mg/L |
| Silver Accuracy | ±0.020 mg/L ±5% of reading |
| Silver Method | Adaptation of the PAN method |
| Sulfate Range | 0 to 150 mg/L (as SO42-) |
| Sulfate Resolution | 1 mg/L |
| Sulfate Accuracy | ±5 mg/L ±3% of reading |
| Sulfate Method | Turbidimetric – Sulfate is precipitated with barium chloride crystals |
| Zinc Range | 0.00 to 3.00 mg/L (as Zn) |
| Zinc Resolution | 0.01 mg/L |
| Zinc Accuracy | ±0.03 mg/L ±3% of reading |
| Zinc Method | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Zincon method |
| Input Channels | 1 pH electrode input and 5 photometer wavelengths |
| pH Electrode | digital pH electrode (not included) |
| Photometer/Colorimeter Light Source | 5 LEDs with 420 nm, 466 nm, 525 nm, 575 nm, and 610 nm narrow band interference filters |
| Photometer/Colorimeter Light Detector | silicon photodetector |
| Bandpass Filter Bandwidth | 8 nm |
| Bandpass Filter Wavelength Accuracy | ±1 nm |
| Cuvette Type | round, 24.6 mm |
| Number of Methods | 128 max. |
| GLP | calibration data for connected pH electrode |
| Display | 128 x 64 pixel LCD with backlight |
| Logging Type | log on demand with user name and sample ID optional input |
| Logging Memory | 1000 readings |
| Connectivity | USB-A host for flash drive; micro-USB-B for power and computer connectivity |
| Power Supply | 5 VDC USB 2.0 power adapter with USB-A to micro-USB-B cable (included) |
| Battery Type/Life | 3.7 VDC Li-polymer rechargeable battery / >500 photometric measurements or 50 hours of continuous pH measurement |
| Environment | 0 to 50.0 ℃ (32 to 122.0 °F); 0 to 95% RH, non-condensing |
| Weight | 1.0 kg (2.2 lbs.) |
| Dimensions | 206 x 177 x 97 mm (8.1 x 7.0 x 3.8”) |